{"id":2427,"date":"2024-08-10T06:44:23","date_gmt":"2024-08-10T06:44:23","guid":{"rendered":"https:\/\/incurelab.com\/wp\/?p=2427"},"modified":"2026-08-27T06:25:23","modified_gmt":"2026-08-27T06:25:23","slug":"incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","status":"publish","type":"post","link":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","title":{"rendered":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly"},"content":{"rendered":"<p>When a lens, a filter, or a photodiode is glued into a benchtop diagnostic instrument, a few micrometers of drift during cure can push the optical path out of specification. The adhesive has to lock the part where the alignment fixture set it and keep it there through temperature changes and shipping.<\/p>\n<h3>The positional-stability problem<\/h3>\n<p>Most adhesives shrink as they cure. Even a small volumetric shrinkage, concentrated in a thin bond line, pulls the bonded part off its aligned position. Then, over the following days, the adhesive continues to relax and creep, adding slow post-cure drift. For a structural joint that does not matter. For an optical mount it is the difference between a passing and a failing unit.<\/p>\n<p>Incure&#8217;s <strong>Cyro-Weld\u2122 5000-series<\/strong> UV adhesives include grades formulated for low linear shrinkage and low post-cure creep, such as <strong>5013<\/strong> and <strong>5017<\/strong>. They cure in seconds under UV or visible light, so the part is fixed at the instant the alignment fixture is still holding it, and the low shrinkage means it stays within a tight positional window afterward. The cured adhesives are formulated to meet ISO 10993-5 and are validated for EtO and Gamma sterilization, which matters for the external diagnostic consumables and reusable instrument optics they are used in.<\/p>\n<h3>Where these adhesives are used<\/h3>\n<ul>\n<li>Mounting lenses, prisms, and windows in optical readers and analyzers<\/li>\n<li>Bonding photodiodes, LEDs, and image sensors to their carriers<\/li>\n<li>Fixing fiber ferrules and collimators in place after active alignment<\/li>\n<li>Sealing and positioning optical filters in fluorescence detection paths<\/li>\n<li>Attaching optical components in handheld and wearable diagnostic modules<\/li>\n<\/ul>\n<p>None of these are implanted; they are external instrument and consumable assemblies where optical precision drives yield.<\/p>\n<h3>Cure without disturbing alignment<\/h3>\n<p>The value of a UV cure here is that it is triggered on command. The operator or the machine aligns the part, confirms the optical reading is in spec, and only then exposes the joint to light. Contrast that with a two-part epoxy, where the bond is drifting throughout a long room-temperature cure while nothing holds the part.<\/p>\n<p>For transparent optical joints, the adhesive&#8217;s own clarity, refractive index, and color stability under UV exposure become part of the optical budget. Incure&#8217;s discussion of <a href=\"https:\/\/incurelab.com\/wp\/uv-glue-vs-epoxy-which-is-better-for-transparent-bonding\">UV adhesives for transparent bonding<\/a> is directly relevant.<\/p>\n<h3>Delivering a controlled dose<\/h3>\n<p>Optical bonds are usually small and often shadowed by the component itself, so a focused spot lamp with a light guide is the common cure tool. See Incure&#8217;s overview of <a href=\"https:\/\/incurelab.com\/wp\/what-is-a-light-guide-in-a-uv-spot-lamp-system\">what a light guide does in a UV spot lamp system<\/a> and <a href=\"https:\/\/incurelab.com\/wp\/incure-l9000-uv-led-spot-lamp-matching-lightguide-to-reach-and-working-distance\">matching a spot-lamp light guide to reach and working distance<\/a>. Grades with a secondary cure mechanism finish crosslinking any resin the spot could not reach.<\/p>\n<h3>Managing thermal drift after assembly<\/h3>\n<p>Even a perfectly aligned, fully cured optical mount can walk out of spec if the adhesive, the mount, and the optic expand at different rates. The bond line acts as a compliant layer; too thick and it allows movement, too thin and it transmits stress into the optic. Incure&#8217;s explanation of <a href=\"https:\/\/incurelab.com\/wp\/how-cte-mismatch-causes-adhesive-bond-failure\">how CTE mismatch causes bond failure<\/a> applies to optical mounts as much as to structural ones. Selecting a grade with the right cured modulus, and controlling bond-line thickness with shims or a fixture, keeps thermal drift inside the optical tolerance.<\/p>\n<h3>Bond-line thickness is a design parameter<\/h3>\n<p>For an optical mount, the thickness of the adhesive layer is not an incidental result of assembly; it is a number to be specified and held. A thicker bond line has more total shrinkage and more compliance, so it allows more initial movement and more thermal drift. A very thin bond line transmits stress straight into a glass optic and can distort a wavefront or, in the worst case, crack the part. Most precision optical mounts target a bond line in the range of 25 to 100 micrometers, set with shims, glass beads mixed into the adhesive, or a hard stop in the fixture.<\/p>\n<p>Three small, symmetric adhesive pads generally hold an optic more stably than one continuous ring, because three points define a plane and the shrinkage of each pad is balanced against the others. An asymmetric bond pattern shrinks unevenly and tips the optic.<\/p>\n<h3>Managing photoinitiator and clarity<\/h3>\n<p>Optical adhesives carry a photoinitiator that absorbs in the near-UV. Any unreacted initiator left after cure can slowly yellow under service light or continue to absorb in the blue, which matters in a fluorescence path. A full cure consumes the initiator; an under-cure leaves it active. This is another reason dose verification at the fixture, not just at the lamp, belongs in the process.<\/p>\n<h3>Qualification<\/h3>\n<p>An optical-bond qualification typically measures initial alignment retention, alignment after thermal cycling, alignment after the sterilization dose, and long-term positional stability under accelerated aging. Locking bond-line thickness, cure dose, and surface preparation at the end of that study is what makes production repeatable.<\/p>\n<h3>Get a recommendation<\/h3>\n<p>Send Incure the optic and mount materials, the positional tolerance, the bond-line geometry, and the sterilization method. <a href=\"mailto:support@incurelab.com\">Email Us<\/a> for a grade and process recommendation.<\/p>\n<p>To evaluate samples or get support qualifying an optical-bonding step, <a href=\"https:\/\/www.incurelab.com\/contact\">Contact Our Team<\/a>.<\/p>\n<p>Visit <a href=\"https:\/\/www.incurelab.com\">www.incurelab.com<\/a> for more information.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>When a lens, a filter, or a photodiode is glued into a benchtop diagnostic instrument, a few micrometers of drift during cure can push the optical path out of specification. The adhesive has to lock the part where the alignment fixture set it and keep it there through temperature changes and shipping. The positional-stability problem [&hellip;]<\/p>\n","protected":false},"author":9,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"ocean_front_end_style_editor":"no","ocean_post_layout":"","ocean_both_sidebars_style":"","ocean_both_sidebars_content_width":0,"ocean_both_sidebars_sidebars_width":0,"ocean_sidebar":"","ocean_second_sidebar":"","ocean_disable_margins":"enable","ocean_add_body_class":"","ocean_shortcode_before_top_bar":"","ocean_shortcode_after_top_bar":"","ocean_shortcode_before_header":"","ocean_shortcode_after_header":"","ocean_has_shortcode":"","ocean_shortcode_after_title":"","ocean_shortcode_before_footer_widgets":"","ocean_shortcode_after_footer_widgets":"","ocean_shortcode_before_footer_bottom":"","ocean_shortcode_after_footer_bottom":"","ocean_display_top_bar":"default","ocean_display_header":"default","ocean_header_style":"","ocean_center_header_left_menu":"","ocean_custom_header_template":"","ocean_custom_logo":0,"ocean_custom_retina_logo":0,"ocean_custom_logo_max_width":0,"ocean_custom_logo_tablet_max_width":0,"ocean_custom_logo_mobile_max_width":0,"ocean_custom_logo_max_height":0,"ocean_custom_logo_tablet_max_height":0,"ocean_custom_logo_mobile_max_height":0,"ocean_header_custom_menu":"","ocean_menu_typo_font_family":"","ocean_menu_typo_font_subset":"","ocean_menu_typo_font_size":0,"ocean_menu_typo_font_size_tablet":0,"ocean_menu_typo_font_size_mobile":0,"ocean_menu_typo_font_size_unit":"px","ocean_menu_typo_font_weight":"","ocean_menu_typo_font_weight_tablet":"","ocean_menu_typo_font_weight_mobile":"","ocean_menu_typo_transform":"","ocean_menu_typo_transform_tablet":"","ocean_menu_typo_transform_mobile":"","ocean_menu_typo_line_height":0,"ocean_menu_typo_line_height_tablet":0,"ocean_menu_typo_line_height_mobile":0,"ocean_menu_typo_line_height_unit":"","ocean_menu_typo_spacing":0,"ocean_menu_typo_spacing_tablet":0,"ocean_menu_typo_spacing_mobile":0,"ocean_menu_typo_spacing_unit":"","ocean_menu_link_color":"","ocean_menu_link_color_hover":"","ocean_menu_link_color_active":"","ocean_menu_link_background":"","ocean_menu_link_hover_background":"","ocean_menu_link_active_background":"","ocean_menu_social_links_bg":"","ocean_menu_social_hover_links_bg":"","ocean_menu_social_links_color":"","ocean_menu_social_hover_links_color":"","ocean_disable_title":"default","ocean_disable_heading":"default","ocean_post_title":"","ocean_post_subheading":"","ocean_post_title_style":"","ocean_post_title_background_color":"","ocean_post_title_background":0,"ocean_post_title_bg_image_position":"","ocean_post_title_bg_image_attachment":"","ocean_post_title_bg_image_repeat":"","ocean_post_title_bg_image_size":"","ocean_post_title_height":0,"ocean_post_title_bg_overlay":0.5,"ocean_post_title_bg_overlay_color":"","ocean_disable_breadcrumbs":"default","ocean_breadcrumbs_color":"","ocean_breadcrumbs_separator_color":"","ocean_breadcrumbs_links_color":"","ocean_breadcrumbs_links_hover_color":"","ocean_display_footer_widgets":"default","ocean_display_footer_bottom":"default","ocean_custom_footer_template":"","ocean_post_oembed":"","ocean_post_self_hosted_media":"","ocean_post_video_embed":"","ocean_link_format":"","ocean_link_format_target":"self","ocean_quote_format":"","ocean_quote_format_link":"post","ocean_gallery_link_images":"on","ocean_gallery_id":[],"footnotes":""},"categories":[103],"tags":[1598,1603,1612,1621,1620,1623,333,1601,1619,1613,1622,1610,1615],"class_list":["post-2427","post","type-post","status-publish","format-standard","hentry","category-uv-adhesives","tag-biocompatible-adhesive","tag-eto-sterilization","tag-gamma-sterilization","tag-high-elongation","tag-incure-cyro-weld-5942","tag-iso-10993-5","tag-medical-adhesive","tag-medical-device-manufacturing","tag-medical-optics","tag-multi-substrate-bonding","tag-precision-bonding","tag-thermal-cycling","tag-uv-visible-light-led-curable-adhesive","entry"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.4 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.\" \/>\n<meta property=\"og:description\" content=\"When a lens, a filter, or a photodiode is glued into a benchtop diagnostic instrument, a few micrometers of drift during cure can push the optical path out of specification. The adhesive has to lock the part where the alignment fixture set it and keep it there through temperature changes and shipping. The positional-stability problem [&hellip;]\" \/>\n<meta property=\"og:url\" content=\"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\" \/>\n<meta property=\"og:site_name\" content=\"INCURE INC.\" \/>\n<meta property=\"article:published_time\" content=\"2024-08-10T06:44:23+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-08-27T06:25:23+00:00\" \/>\n<meta name=\"author\" content=\"Incure Content Team\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Written by\" \/>\n\t<meta name=\"twitter:data1\" content=\"Incure Content Team\" \/>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\"},\"author\":{\"name\":\"Incure Content Team\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#\\\/schema\\\/person\\\/53482b163015e8f48dc86f4132515727\"},\"headline\":\"UV Adhesives for Optical Alignment in Diagnostic Device Assembly\",\"datePublished\":\"2024-08-10T06:44:23+00:00\",\"dateModified\":\"2026-08-27T06:25:23+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\"},\"wordCount\":862,\"publisher\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#organization\"},\"keywords\":[\"biocompatible adhesive\",\"EtO sterilization\",\"gamma sterilization\",\"high elongation\",\"Incure Cyro-Weld\u2122 5942\",\"ISO 10993-5\",\"medical adhesive\",\"medical device manufacturing\",\"medical optics\",\"multi-substrate bonding\",\"precision bonding\",\"thermal cycling\",\"UV\\\/Visible Light\\\/LED curable adhesive\"],\"articleSection\":[\"UV Adhesives\"],\"inLanguage\":\"en-US\"},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\",\"url\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\",\"name\":\"UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#website\"},\"datePublished\":\"2024-08-10T06:44:23+00:00\",\"dateModified\":\"2026-08-27T06:25:23+00:00\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive\"]}]},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"UV Adhesives for Optical Alignment in Diagnostic Device Assembly\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#website\",\"url\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/\",\"name\":\"INCURE INC.\",\"description\":\"Engineered Solutions. Trusted Worldwide\",\"publisher\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#organization\"},\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"},{\"@type\":\"Organization\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#organization\",\"name\":\"INCURE INC.\",\"url\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/\",\"logo\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#\\\/schema\\\/logo\\\/image\\\/\",\"url\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/wp-content\\\/uploads\\\/2025\\\/07\\\/incure-logo.jpeg\",\"contentUrl\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/wp-content\\\/uploads\\\/2025\\\/07\\\/incure-logo.jpeg\",\"width\":164,\"height\":45,\"caption\":\"INCURE INC.\"},\"image\":{\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#\\\/schema\\\/logo\\\/image\\\/\"}},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/#\\\/schema\\\/person\\\/53482b163015e8f48dc86f4132515727\",\"name\":\"Incure Content Team\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g\",\"caption\":\"Incure Content Team\"},\"url\":\"https:\\\/\\\/incurelab.com\\\/wp\\\/author\\\/kimleelcy\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","og_locale":"en_US","og_type":"article","og_title":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.","og_description":"When a lens, a filter, or a photodiode is glued into a benchtop diagnostic instrument, a few micrometers of drift during cure can push the optical path out of specification. The adhesive has to lock the part where the alignment fixture set it and keep it there through temperature changes and shipping. The positional-stability problem [&hellip;]","og_url":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","og_site_name":"INCURE INC.","article_published_time":"2024-08-10T06:44:23+00:00","article_modified_time":"2026-08-27T06:25:23+00:00","author":"Incure Content Team","twitter_card":"summary_large_image","twitter_misc":{"Written by":"Incure Content Team","Est. reading time":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#article","isPartOf":{"@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive"},"author":{"name":"Incure Content Team","@id":"https:\/\/incurelab.com\/wp\/#\/schema\/person\/53482b163015e8f48dc86f4132515727"},"headline":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly","datePublished":"2024-08-10T06:44:23+00:00","dateModified":"2026-08-27T06:25:23+00:00","mainEntityOfPage":{"@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive"},"wordCount":862,"publisher":{"@id":"https:\/\/incurelab.com\/wp\/#organization"},"keywords":["biocompatible adhesive","EtO sterilization","gamma sterilization","high elongation","Incure Cyro-Weld\u2122 5942","ISO 10993-5","medical adhesive","medical device manufacturing","medical optics","multi-substrate bonding","precision bonding","thermal cycling","UV\/Visible Light\/LED curable adhesive"],"articleSection":["UV Adhesives"],"inLanguage":"en-US"},{"@type":"WebPage","@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","url":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive","name":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly - INCURE INC.","isPartOf":{"@id":"https:\/\/incurelab.com\/wp\/#website"},"datePublished":"2024-08-10T06:44:23+00:00","dateModified":"2026-08-27T06:25:23+00:00","breadcrumb":{"@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive"]}]},{"@type":"BreadcrumbList","@id":"https:\/\/incurelab.com\/wp\/incure-cyro-weld-5942-uv-curable-pressure-sensitive-adhesive#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/incurelab.com\/wp\/"},{"@type":"ListItem","position":2,"name":"UV Adhesives for Optical Alignment in Diagnostic Device Assembly"}]},{"@type":"WebSite","@id":"https:\/\/incurelab.com\/wp\/#website","url":"https:\/\/incurelab.com\/wp\/","name":"INCURE INC.","description":"Engineered Solutions. Trusted Worldwide","publisher":{"@id":"https:\/\/incurelab.com\/wp\/#organization"},"potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/incurelab.com\/wp\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"},{"@type":"Organization","@id":"https:\/\/incurelab.com\/wp\/#organization","name":"INCURE INC.","url":"https:\/\/incurelab.com\/wp\/","logo":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/incurelab.com\/wp\/#\/schema\/logo\/image\/","url":"https:\/\/incurelab.com\/wp\/wp-content\/uploads\/2025\/07\/incure-logo.jpeg","contentUrl":"https:\/\/incurelab.com\/wp\/wp-content\/uploads\/2025\/07\/incure-logo.jpeg","width":164,"height":45,"caption":"INCURE INC."},"image":{"@id":"https:\/\/incurelab.com\/wp\/#\/schema\/logo\/image\/"}},{"@type":"Person","@id":"https:\/\/incurelab.com\/wp\/#\/schema\/person\/53482b163015e8f48dc86f4132515727","name":"Incure Content Team","image":{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/secure.gravatar.com\/avatar\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/85e856c6e17c9c44477ea1e0964ef5dc19c2a1ef29a741bb9bb742fcfc644fcc?s=96&d=mm&r=g","caption":"Incure Content Team"},"url":"https:\/\/incurelab.com\/wp\/author\/kimleelcy"}]}},"rttpg_featured_image_url":null,"rttpg_author":{"display_name":"Incure Content Team","author_link":"https:\/\/incurelab.com\/wp\/author\/kimleelcy"},"rttpg_comment":0,"rttpg_category":"<a href=\"https:\/\/incurelab.com\/wp\/category\/uv-adhesives\" rel=\"category tag\">UV Adhesives<\/a>","rttpg_excerpt":"When a lens, a filter, or a photodiode is glued into a benchtop diagnostic instrument, a few micrometers of drift during cure can push the optical path out of specification. The adhesive has to lock the part where the alignment fixture set it and keep it there through temperature changes and shipping. The positional-stability problem&hellip;","_links":{"self":[{"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/posts\/2427","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/users\/9"}],"replies":[{"embeddable":true,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/comments?post=2427"}],"version-history":[{"count":17,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/posts\/2427\/revisions"}],"predecessor-version":[{"id":22105,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/posts\/2427\/revisions\/22105"}],"wp:attachment":[{"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/media?parent=2427"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/categories?post=2427"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/incurelab.com\/wp\/wp-json\/wp\/v2\/tags?post=2427"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}